FortisCare EMMU with Bluetooth Dongle
|
FortisCare EMMU w/
Bluetooth Dongle
| Meter Specifications |
| Memory capacity |
480 results (glucose level, date & time) |
| Test specimen |
Capillary whole blood / Venous whole blood |
| Sample volume |
0.6 ul |
| Test time |
5 seconds |
| AST |
Fingertip, Palm, Forearm |
| HCT range |
20-60% |
| Measuring range |
20~600 mg/dL |
| Operating ranges |
10°C~40°C, below ≦90% R.H. |
| How calibrated |
No code |
| Auto-shut off |
Power Off functions after 90 seconds |
| Meter dimension |
96.52 x 50.8 x 15.24 mm |
| Weight |
39.12g (without battery) |
| Battery life |
Over 2,000 times |
| Power supply |
Alkaline Battery 1.5V x2 (AAA) |
| Time Setting |
12 Hours or 24 Hours (Optional) |
| PC Link |
USB port for data download |
| Principles |
Electrochemical biosensor technology |
| Standards |
IVD 98/79/EC
ISO 13485 : 2003
ISO 14971 : 2007
ISO 15197 : 2003
ISO 15223 : 2000
EN 980 : 2003
EN 60601-1 : 1990
EN 60601-1-2 : 2002
EN 61010-2-101 : 2002
|
| Error Codes |
Err: humidified / Used strips
LP: Low power, please replace new battery
001: System Fault: Not enough blood
005: System Fault
HI: test result is higher than 630 mg/dL
LO: test result is lower than 20 mg/dL
Ht: Temperature too high to have a test
Lt: Temperature too low to have a test
|
|
|
|