FortisCare EM66

FortisCare

Features
  • No coding system
  • Meets ISO 15197 accuracy requirement
  • Free from 61 substances
  • Under-fill blood volume detection
  • Ketone warning
  • Before/after meal marking
  • Inverse backlight display

FortisCare EM66

Blood Glucose Meter

Meter Specifications
Blood specimen Capillary whole blood/ Venous whole blood
Memory capacity 480
Blood sample 0.6 μL Fresh blood from fingertip, palm, or forearm
Standards IVD 98/79/EC
ISO 13485 : 2003
ISO 14971 : 2009
ISO 15197 : 2013
EN 980 : 2008
EN 61326-1 : 2006
EN 61326-2-6 : 2005
EN 61010-2-101 : 2002
Test time 5 seconds
AST Palm and Forearm
HCT range 25~55%
Measuring range 20~600 mg/dL
Accuracy 95% within + 15 mg/dL when glucose <100mg/dL 95% within + 15% when glucose >100mg/dL
Precision CV (correlation variation) is less than 5 %
Storage & Transportation Temp. 35.6°F -86°F (2°C -30°C)
Temperature and Humidity Operating : 10℃~40℃, below 85% R.H.
Meter storage : -10℃~65℃, below 90% R.H.
Strip storage : 2℃~30℃, below 90% R.H.
Transportation : 2℃~30℃, below 90% R.H.
How calibrated No code
Meter dimension 83.7 x 51.2 x 22.9mm
Power Supply Alkaline Battery 1.5v x2 (AAA)
Battery life Over 1000 tests
Data port USB
Auto shut-off Power off functions after 90 seconds no key presse
Time Setting 12 Hours or 24 Hours (Optional)
Principles Electrochemical biosensor technology
Display Liquid crystal display module With inverse backlight Dimension : 56.8 x 37 mm
Error Codes Err : humidified / Used strips
LP : Low power, please replace new battery
001 with err icon : Not enough blood
005 with err icon : system fault
HI : test result is higher than 630 mg/dL
LO: test result is lower than 20 mg/dL
Ht : Temperature too high to have a test
Lt : Temperature too low to have a test